Installation/assembly – Yamato Scientific RE540 Evaporators, Rotary User Manual
Page 12

9
Installation/Assembly
3. Inserting the steam duct into the motor
1.Before you insert the steam duct, make sure the O-rings inside the motor center hole are not out
of place. If so, reset them in the right place.
2.Insert the steam duct from the right side into the center hole of the motor.
3.Set the steam duct to the desired position with a minimum of 5mm between the blue flask
remover and the steam duct holder.
4.Tighten the steam duct holder by turning clockwise. Be sure to tighten the steam duct holder
firmly so that the steam duct does not slip.
※
※
※
※When removing the steam duct, first loosen the steam duct holder. Do not remove the
steam duct holder or the rings may slip out.
●
●
●
●If the rings come off in setting / removing the steam duct and you do not know how to
assemble…..
See the following picture to re-assemble.
Steam Duct Holder
Flask Remover
Keep at least 5mm
Rubber O- Ring