10

Installation/Assembly

4. Set the cooling condenser nut, coil ring and vacuum seal to the condenser or

condenser connector.

1. Remove the cooling condenser nut and coil ring on the left side of motor.

2. Connect the cooling condenser nut and then the coil ring to the condenser or condenser

connector (which ever applies). Be sure to put the nut on first.

3.Insert the vacuum seal to the condenser or condenser connector (which ever applies).

※ It is optional to put grease on the vacuum seal.

Insert the convex side of the vacuum

seal into the flange of cooling

condenser or condenser connector.

5. Connect cooling condenser or cooling condenser connector to the motor.

Insert the steam duct into vacuum seal, put glass flange to the motor and fasten firmly the cooling

condenser nut.

Be careful not to damage the set vacuum seal when you put steam duct into the flange

with the seal in case the damage could cause leak.

Glass Type A

Glass Type B/C

Above :

Fitting of Coil Ring and Vacuum Seal

Below :

Fitting to Motor

Above :

Fitting of Coil Ring and Vacuum Seal

Below :

Fitting to Motor

When you remove the coil ring from cooling condenser or condenser connector

Hook the coil ring by the attached hexagon wrench as shown in the left

picture in order to remove the ring easily from the cooling condenser or

connecting condenser. However, be careful not to force too hard in case

the glass apparatus or the coil ring might be damaged.