Installation/assembly, Assembling parts – Yamato Scientific RE550 Evaporators, Rotary User Manual
Page 10

8
Installation/Assembly
Assembling parts
3. Inserting the steam duct into the motor
1.Before installing the steam duct check that the O-rings are in the center hole of motor.
If not, be sure to set them in the correct place (See figure 1 below).
2.Insert the steam duct from the right side into the center hole of the motor.
3.Set the steam duct to the desired position with a minimum of 5mm between the blue flask
remover and the steam duct holder.
4.Tighten the steam duct holder by turning clockwise. Be sure to tighten the steam duct holder
firmly so the steam duct does not slip.
※
※
※
※When removing the steam duct, first loosen the steam duct holder. Do not remove the
steam duct holder or the rings may slip out.
●
●
●
●If the rings come off in setting / removing the steam duct, please refer to figure 1.
O-ring
Steam Duct Holder
Flask Remover
Keep at least 5mm
Rubber O- Ring
Figure 1