Installation/Assembly

Assembling parts

4. Set the cooling condenser nut, coil ring and vacuum seal to the condenser or

condenser connector.

1. Remove the cooling condenser nut and coil ring on the left side of motor.

2. Connect the cooling condenser nut and then the coil ring to the condenser or condenser

connector (which ever applies). Be sure to put the nut on first.

3.Insert the vacuum seal to the condenser or condenser connector (which ever applies).

※ It is optional to put grease on the vacuum seal.

Insert the convex side of the vacuum

seal into the flange of cooling

condenser or condenser connector.

Glass Type A

Glass Type B/C

5. Connect cooling condenser or cooling condenser connector to the motor.

Insert the steam duct into vacuum seal, put glass flange to the motor and fasten firmly the cooling

condenser nut.

Be careful not to damage the glass coil in cooling condenser when inserting the steam duct.

Be careful not to damage the set vacuum seal when inserting the steam duct.

Glass Type A

Glass Type B/C

Removing the coil ring from cooling condenser or condenser connector

Hook the coil ring by the attached hexagon wrench as shown in the left

picture in order to remove the ring easily from the cooling condenser or

connecting condenser.

Apply moderate force to remove the coil ring; extreme force may damage

the coil ring and or the glass apparatus.

Coil Ring

Vacuum Seal

Coil Ring

Vacuum Seal

Vacuum Seal

Cooling Condenser Nut

Coil Ring