Troubleshooting Guide

Problem Solving Chart

Trouble & Countermeasure

Check the following points if trouble occurs.

Contact Yamato’s Technical Service Department for further information.

Trouble

Cause

Countermeasure

Digital display does not

light up on the controller

● Power is off

● Disconnection of power cord

● Blown fuse

● Check power source

● Connect the main power cord

● Connect motor power cord

● Exchange fuse (2A)

Lift will not move

● Power is off

● Disconnection of power cord

● Fuse blows

Overload (any non-standard

heavier glass set or apparatus

is equipped )

● Turn the power on

● Connect the cord of body

● Exchange of fuse (2A)

● Provide the specified glass set or

apparatus

The flask will not rotate

● Switch of controller is off

● Control knob is at “min”

● Disconnection of motor cable

● Incomplete set-up or fastening

of steam duct cause racing

● Something touching the flask

● Turn on the power switch

● Turn the control knob up

● Insert into the socket on the

controller

● Fasten the steam duct holder

●Remove obstruction

Incomplete vacuum

● Wear and deterioration of

vacuum seal

● Direction of vacuum seal is

wrong

● Cooling condenser nut is

incompletely fastened

● Glass apparatus break

● Incomplete connection of glass

apparatus

● Leak from hose joints

● Exchange of vacuum seal

● Re-set the vacuum seal

● Re-fasten

● Exchange

● Re-set

● Put vacuum grease on

●Check, re-fasten and put vacuum

grease on joints