6 troubleshooting, 7 technical specifications – Metrohm 940 Professional IC Vario ONE/LPG User Manual

Page 84

Advertising
background image

6.1

■■■■■■■■■■■■■■■■■■■■■■

76

■■■■■■■■

940 Professional IC Vario ONE/LPG (2.940.1150)

6 Troubleshooting

Problem

Cause

Remedy

Marked drop in
pressure.

Leak in the system.

Check all capillary connections and seal leaks,
if necessary (see Chapter 3.2, page 14).

The baseline has a
large amount of
noise.

The eluent is not suffi-
ciently degassed.

Ensure that the connectors for the eluent
degasser are firmly connected (see Chapter
5.6, page 54)
.

The eluent path has a leak.

Check the eluent path and fix the leak.

High-pressure pump – Con-
taminated pump valves.

Clean pump valves (see Chapter 5.8, page 56).

The eluent path is blocked.

Check the eluent path and fix the blockage.

High-pressure pump –
Defective piston seals.

Replace the piston seals (see Chapter 5.8,
page 56)
.

The pulsation absorber is
not connected or is defec-
tive.

Connect (see Chapter 3.11, page 27) or
replace the pulsation absorber.

The baseline is drift-
ing.

Thermal equilibrium is not
yet attained.

Condition the instrument with the column
thermostat enabled until the baseline is stable .

Leak in the system.

Check all capillary connections and seal leaks,
if necessary (see Chapter 3.2, page 14).

The organic solvent in the
eluent is evaporating.

■

Check the eluent bottle cap (see Chapter
3.7, page 22)
.

■

Constantly stir the eluent.

The pressure in the
system markedly
increases.

The inline filter
(6.2821.120) is blocked.

Replace the filter (6.2821.130) .

Guard column – Blocked.

Replace the guard column (see Chapter 3.20,
page 41)
.

Separation column –
Blocked.

■

Regenerate the separation column (see
Chapter 5.14.4, page 74)
.

■

Replace the separation column (see "Con-
necting the separation column", page 45)
.

Note: Samples should always be microfiltered .