Metrohm 748 DH Sample Changer User Manual

Page 16

Advertising
background image

3.2 Remote control commands

748 DH Sample Changer

14

Ø

- Stirrs during 3 s
- Moves up to Z=0, SpeedZFast
.Times: The number of washing cycles can be set. If the number is >1, the
following additional steps will be carried out (n times); the repetitive steps be-
ginning at the marked position Ø in the sequence:
- Moves up to Z=0, SpeedZFast
- Switches pump on
- Moves down to Zmax., Speed Zslow
- Switches pump off

3.2.2.7

SetCtrlLine

$G

SetCtrlLine.On

True, False

SetCtrlLine.Line

1...3

A remote control line can be set. Select socket 1...3 and set line as follows:
&SetCtrlLine.On"True";..$G Sets an active line (0 V), see also page 20.
&SetCtrlLine.On"False";..$G Sets the line to inactive (5 V)
The beaker sensor is normally plugged into socket 1.

3.2.2.8

WaitForInput

$G

WaitForInput.On

True, False

WaitForInput.Line

1...3

The Sample Changer waits for an active line (with WaitForInput.ON"True") or
for an inactive line (with WaitForInput.ON"False"). No other commands are
worked off during this waiting time.

3.2.2.9

Init

$G

Init.SpeedXY

0...5...7

Init.SpeedXYFast

0...2...7

Init.SpeedZ

0...3...7

Init.SpeedZFast

0...7

Init.SpeedZSlow

0...7

Sets the global speeds for all makro commands of the Sample Changer. A
small number means a high speed, i.e. 0 is the highest possible speed.
The speeds are only set if the init movement is carried out (command &Init
$G).
&Init $G moves to home position and Z=0.

3.2.2.10 StopAll

$G

Emergency stop. Stops all commands immediatelay. May be used e.g. to set
remote control lines to inactive, or to stop stirrer or pump.