Metrohm 717 Sample Changer User Manual
Page 24

5.1 Technical Specifications
717 Sample Changer, Appendix
22
5.1.2 759 Swing Head
Dimensions
Height: 0.14 m, Width: 0.10 m, Depth: 0.09 m
Weight
0.63 kg
Material
Swing Head case:
Polybuteneterephthalate (PBTP)
with metallic fiber
Titration Head:
Polypropylene
Rotation
4 fix positions
Temperatures
Nominal functional range 5 – 40 °C
Transport and storage –40 – +70 °C
Distribution
5 V DC, drawing of current 500 mA
voltage
connection to the remote socket of the Metrohm Sample
Changers, e.g. model 717, 730, 760, ...
All data are typical values.
Safety Specifications
Constructed and tested according to IEC 1010 / EN 61010 /
UL 3101-1, Safety Class III
Degree of protection IP 43
Electromagnetic Compatibility (EMC)
Emitted Interference: The instrument complies with the basic specifications
EN 50081-1/2 1992, EN 55011 (class B), EN 55022
(class B), and NAMUR.
Interference immunity: The basic specifications EN 50082-1 1997, and IEC 801-2 to
IEC 801-6 are adhered to.