4. Appendix

700 Dosino, Instructions for Use

39

(

)

1

2

−

−

=

∑

n

V

V

s

i

s : repeatability standard deviation
V

i

: distributed volume

n : number of distribution

Please note

•

The above-mentioned limits are related to water at 20 °C and
are valid for dosing units with glass cylinders only.

•

The nominal volume of the dosing cylinders are printed on the
housing of the dosing units.

•

The above-mentioned limits are valid for the useful volume ran-
ge of 10 % to 100 % of the nominal volume.

•

The error limits relate to the ISO 8655-3 Draft International
Standard (Piston-operated volumetric apparatus - part 3: Piston
burettes), which is to come to effect in the year 2000.

•

Metrohm burets with glass cylinders in general have a better
quality performance (systematic and random error) than the
maximum error limits implied. Check your dosing system ac-
cording to the ISO test specification.

Possible sources of dosing errors

Influencing parameters are:

•

Air bubbles in the dosing cylinder or tubings.

•

Leaky piston/cylinder system.

•

Leaky tubing connections.

•

Differences in hydrostatic pressure of the liquid level in the
supply flask versus the dosing tip.

•

Viscosity and/or flow characteristics of the liquid to be dosed.