6 Troubleshooting

930 Compact IC Flex ChS (2.930.1200)

■■■■■■■■

81

6 Troubleshooting

Problem

Cause

Remedy

Marked drop in
pressure.

Leak in the system.

Check all capillary connections and seal leaks,
if necessary (see Chapter 3.2, page 12).

The baseline has a
large amount of
noise.

The eluent path has a leak.

Check the eluent path and fix the leak.

High-pressure pump – Con-
taminated pump valves.

Clean pump valves (see Chapter 5.6, page 53).

The eluent path is blocked.

Check the eluent path and fix the blockage.

High-pressure pump –
Defective piston seals.

Replace the piston seals (see Chapter 5.6,
page 53)
.

The pulsation absorber is
not connected or is defec-
tive.

Connect (see Chapter 3.8, page 24) or replace
the pulsation absorber.

The baseline is drift-
ing.

Leak in the system.

Check all capillary connections and seal leaks,
if necessary (see Chapter 3.2, page 12).

The organic solvent in the
eluent is evaporating.

■

Check the eluent bottle cap (see Chapter
3.5, page 18)
.

■

Constantly stir the eluent.

The pressure in the
system markedly
increases.

The inline filter
(6.2821.120) is blocked.

Replace the filter (6.2821.130) .

Suppressor blocked.

■

Regenerate the suppressor (see Chapter
5.10.3.2, page 71)
.

Note: 6.2821.180 pump tubing connection
with filter must be used.

Guard column – Blocked.

Replace the guard column (see Chapter 3.15,
page 40)
.

Separation column –
Blocked.

■

Regenerate the separation column (see
Chapter 5.13.4, page 79)
.

■

Replace the separation column (see "Con-
necting the separation column", page 44)
.

Note: Samples should always be microfiltered .