103
7.
Push down the fuse door firmly until it
snaps into place. Be sure this has been
performed for both fuses.
8.
Close and latch the door of the XDS Process Analytics Instrument.
9.
Restore power to the instrument.
10.
Connect to the instrument from Vision. Allow time for Purge System activation (if so
equipped) plus about 20 minutes for temperature stabilization.
11.
Run Performance Test to verify instrument operation. Because no optical components were
changed, there is no requirement to run other tests.