Metrohm TitrIC pro II User Manual

Page 14

Advertising
background image

Application Bulletin 354

Installation Instruction of TitrIC pro II

Page 14 of 14

TitrIC pro II_conductivity calibration:

This method is used

for calibration of the conductivity sensor. It has to be
used in combination with the MagIC Net method

TitrIC

pro II_calibration conductivity

.

Import: Click on

Method

at the left (in MagIC Net and

tiamo

TM

), then select under

File / Method manager… / Edit /

Import

Import the methods related to TitrIC pro II with the

MiPT kit from the MagIC Net CD. You will find the methods
in the folder example/TitrIC of the CD.

Note!

Alterations of these methods should only be carried out by a
person at the administrator level and who is thoroughly
familiar with tiamo

TM

and MagIC Net

With the TitrIC pro – Combo Kit Partial Loop setup the
following analysis can be carried out using the TitrIC pro
II_MiPT methods in MagIC respectively tiamo

TM

:

The whole procedure is controlled by the MagIC Net
software which is the master software. In tiamo

TM

the

method

TitrIC pro II_MiPT

is selected as a single

determination. Start the determination in autorepeat mode.
To set it into autorepeat go to the Run window, right-mouse-
click and select

Properties Run window

. Under Process

select

Autorepeat determination

. Close the window with

<ok>

. Start the determination. Now tiamo

TM

is waiting for the

start signal from MagIC Net via RS232.

In the MagIC Net software select the method

TitrIC pro

II_MiPT

. Enter the required sample position, sample

identification etc. into MagIC Net. For value 1 choose
according to the list below:

Value 1 = 1 complete analysis including titration and
ion chromatography.

Value 1 = 2 only an ion chromatography analysis (e.g.
a calibration)

Value 1 = 3 only a titration analysis