3.21 Connecting the separation column

■■■■■■■■■■■■■■■■■■■■■■

48

■■■■■■■■

940 Professional IC Vario ONE/ChS/HPG (2.940.1240)

Start manual control in MagIC Net and select the high-pressure
pump: Manual

Manual control Pump

Flow: in accordance with column leaflet
On

Rinse the guard column with eluent for approx. 5 minutes.

Use manual control in MagIC Net to stop the high-pressure pump
again: Off.

3.21

Connecting the separation column

The intelligent separation column (iColumn) is the heart of the ion chro-
matographic analysis. It separates different components according to their
interactions with the column. Metrohm separation columns are equipped
with a chip where their technical specifications and history (start-up, oper-
ating hours, etc) are stored.

NOTE

Information regarding which separation column is suitable for your
application can be found in the Metrohm IC Column Program, the
product information for your separation column at

http://

www.metrohm.com

(Ion Chromatography product area) or obtained

directly from your representative.

You can find the separation columns and guard columns currently avail-
able from Metrohm in the Metrohm IC Column Program or on the Inter-
net at

http://www.metrohm.com

in the Ion Chromatography product

area. A test chromatogram and a leaflet are provided along with each col-
umn. You can find detailed information on special IC applications in the
corresponding "Application Bulletins" or "Application Notes", which
are available on the Internet at

http://www.metrohm.com

in the Applica-

tions area or free of charge upon request from your responsible Metrohm
representative.

CAUTION

New separation columns are filled with solution and sealed with stop-
pers on both sides. Before inserting the column, ensure that this solu-
tion can be mixed with the eluent being used (follow information from
the manufacturer).