Specifications

Overall Analyzer PRO instrument specifications are as follows:

Item

Specification

Ambient Temperature

23 to 104° F (-5 to 40° C)
With Air Cooling: 23 to 149° F (-5 to 65° C)

Ambient Humidity

10-90% Relative Humidity, non-condensing

Electrical Supply

100-240 VAC, 50-60 Hz, 2.0 A, 150 W

Protection

IP69k according to IEC 60529 and DIN 40050 part 9, NT ELEC 023

Dimensions

Width: 17.0” (42 cm)
Height: 17.0” (42 cm)
Depth: 5.0” (13 cm)

Weight

33.0 pounds (15 Kg)

Pressurized Air Cooling
(Ambient Temperature 45-
65 Degrees C)

Cooling air flow rate minimum 5 Liters/min
>99.9% water free
>99.9% free of oil and fine particles down to 0.3 micrometers

Pressurized Air – Spoon
Probe

Purge air: 3-5 bar

Network Connections

Local Area Network (LAN) – Ethernet, RJ-45 connection

Sample Temperature,
Reflectance Window

302 degrees F (150° C)

Sample Temperature,
Spoon Probe

248 degrees F (120° C)

Computer Requirements

See Vision computer requirements

Wavelength Range

1100-1650 nm

Analysis Time:

5 – 50 ms / integration time; depends upon sample. Typical result
time: 3-15 sec.

Wavelength Accuracy

0.5 nm

Wavelength Precision

< 0.02 nm

Wavelength Stability

< 0.01 nm/°C

Specifications for each type of sample interface follow: