Figure 15, Mounting the tubings – Metrohm 860 KF Thermoprep User Manual
Page 25

■■■■■■■■■■■■■■■■■■■■■■
3 Installation
860 KF Thermoprep
■■■■■■■■
17
2 Screw one 6.1821.040 filter tube into each of the 6.1602.145
drying flask covers from below. Tighten the filter tubes well by
hand.
3 Screw the two drying flask covers with the filter tubes onto the dry-
ing flasks. Tighten the covers well by hand.
Note
If drying flask covers or filter tubes are not sufficiently tightly
screwed on, then this will prevent a precise, regular flow of gas.
The error message "Flow rate error" will appear as a rule when
there are leaks in the threaded connections.
1
1
3
2
4
6.1805.080
6.1805.080
6.1805.010
2
3
Figure 15
Mounting the tubings
Mount the FEP tubings as follows:
1 Load the two drying flasks that have been prepared into the holders,
see illustration.
2 Screw one 6.1805.080 FEP tubing (25 cm length) to the gas outlet
(at the front, labeled 1). Screw the other tubing end on the front dry-
ing flask to the M6 connector without point marking (see left-hand
arrow).