3.2. Remote control commands

798 MPT Titrino

102

window: Only EP's that lie within set windows are recognized.

OFF: The EP evaluation is switched off.

.Recognition.Window.1.LowLim: Lower limit for window in pH, mV (with U

and Ipol) or µA (with Upol).

.Recognition.Window.1.UpLim: Upper limit for window in pH, mV (with U and

Ipol) or µA (with Upol).

Windows are opened until the lower limit is set to OFF. For every

expected EP, an individual window must be set, see page 22.

.FixEP.1.Val: Fix-EP's in pH, mV (for U, Ipol) resp. µA (for Upol). Fix EP's are

evaluated until the setting OFF is found.

.pK: pK or HNP evaluation. Possible only in pH and U titrations.

3.2.2.25. Mode.Parameter.Presel.IReq

id1, id1&2, all, OFF

Mode.Parameter.Presel.SReq

value, unit, all, OFF

Parameters for DET, MET, SET, MEAS: Automatic inquiry after the start of the

determination. From such an inquiry, the determination continues if the re-

quested entry/entries is/are made, e.g. &SmplData.OFFSilo.Id1 (see 3.2.2.84)

or with &M $G, see 3.2.2.1.

$H is not possible in requests.

3.2.2.26. Mode.Parameter.Presel.LimSmplSize.Status ON,

OFF

Mode.Parameter.Presel.LimSmplSize.LoLim

0.0...999 999

Mode.Parameter.Presel.LimSmplSize.UpLim 0.0...999 999

Limit control for the sample size.

3.2.2.27. Mode.Parameter.Presel.ActPuls ON, OFF

for SET:

first, all, cond., OFF

Output of a pulse on the I/O line "Activate", see page 167.


3.2.2.28. Mode.Parameter.SET1.EP pH:

0...±20.00, OFF

U,

Ipol:

0...±2000, OFF

Upol:

0...±200.0, OFF

Mode.Parameter.SET1.UnitEp read

only

Parameters for SET: Setting the 1st endpoint as pH value, in mV (with U and

Ipol) resp. µA (with Upol). The corresponding unit can be read with .UnitEP.

If the value is on "OFF", no further nodes will appear from SET1.

3.2.2.29. Mode.Parameter.SET1.Dyn pH:

0.01...20.00, OFF

U,

Ipol:

1...2000, OFF

Upol:

0.1...200.0, OFF

Mode.Parameter.SET1.UnitDyn read

only

Mode.Parameter.SET1.MaxRate 0.01...10...150, max.

Mode.Parameter.SET1.MinRate 0.01...25.0...999.9

Parameters for SET: Control parameters, see page 30.

.Dyn:

Dynamics, control range in pH, mV (with U and Ipol) or µA (with

Upol). The corresponding unit can be read with .UnitDyn.

.MaxRate: Maximum allowed titration rate in mL/min. Max. means maximum

possible rate with the Exchange Unit in current use.