background image

4 Short operating tutorial

826/827 pH meter, Instructions for Use

17

4

Short operating tutorial

In this section the necessary steps for carrying out a simple pH and ion

measurement with calibration are described. The instructions are li-

mited to those steps that are absolutely necessary and will enable you

to carry out your first measurements with the pH meter directly. The op-

erating principles are described in Section 3.3.

4.1

Requirements

The following instruments, accessories and solutions are required for

carrying out the pH calibration and measurement described below:

•

826 pH mobile (2.826.0XX0) or 827 pH lab (2.827.011X,

2.827.021X)

•

pH electrode

•

Calibration buffers

Metrohm buffer solutions pH 4.00 and pH 7.00

The calibration parameters for the pH mode are set for a calibration us-

ing two Metrohm buffers as default (see Section 6.1.2). If you want to

use other buffers then the corresponding buffer type must be entered.

4.2

pH calibration

1 Start calibration with first buffer

•

Start calibration with <CAL>

•

Immerse pH electrode in buffer solution pH 7 and confirm with

<OK>

•

If a temperature sensor is connected, the calibration tempera-

ture is being determined

•

If no temperature sensor is connected:

enter the temperature with the arrow keys and confirm with

<OK>

•

The first buffer is being measured

•

The following message is displayed after successful mea-

surement:

change buffer <OK>

2 Continue calibration with second buffer

•

Remove pH electrode from first buffer and rinse with water

•

Immerse pH electrode in second buffer solution pH 4 and

continue calibration sequence with <OK>

•

The second buffer is being measured