8.1 Troubleshooting

44

826/827 pH meter, Instructions for Use

Problem

Possible cause

Remedy

Measuring signal absent
or very variable

Electrode not connected

•

Connect electrode

Air in or in front of the diaph-
ragm

•

Remove air

Faulty electrode

•

Replace electrode

Measured value remains
unstable and does not
fulfill drift criterion

Dirty glass membrane or diaph-
ragm

•

Clean membrane or diaphragm

pH or temperature of the solu-
tion is not stable

•

Measure under exclusion of air

•

Thermostat the solution

Unsuitable electrode:

•

Conductivity too low

•

Organic solution

•

Use suitable electrode

Electrode not connected

•

Connect electrode

Faulty electrode

•

Replace electrode

Slow adjustment of
measured value

Dirty glass membrane or diaph-
ragm

•

Clean membrane or diaphragm

Slope too small on cali-
bration

Dirty glass membrane or diaph-
ragm

•

Clean membrane or diaphragm

Glass membrane deswollen af-
ter measurement in non-
aqueous solution

•

Immerse electrode in water between

measurements

Poor buffer solutions

•

Replace buffers

Old Electrode

•

Replace electrode

Measured value obvious-
ly incorrect

Incorrect calibration

•

Check / Repeat calibration

•

Check / Replace buffers

Manual temperature input incor-
rect

•

Enter correct temperature of the solu-

tion

Dirty glass membrane or diaph-
ragm

•

Clean membrane or diaphragm

Electrolyte or electrode too old

•

Refill electrolyte or replace electrode

Faulty electrode

•

Replace electrode