6.1 pH measurement (pH mode)

32

826/827 pH meter, Instructions for Use

The other stored buffer tables are described in Section 9.3. During the

calibration the pH meter shows the type and temperature-specific pH

value of the recognized buffer. For temperatures within the 5 °C steps

the pH value is linearly interpolated.

buff.type:

special

If you would like to use different buffers from those described,

then you require the exact pH value of each buffer at the tempera-

ture used for calibration. These pH values can be entered here.

During the calibration they can again be adapted before each

measurement.

A general rule is that you can only calibrate in defined temperature

ranges, as otherwise the following error message appears:

buffer not defined

6.1.3

Limits

param

:

├

cal.limits

: ├

lo.lim.slope

├

up lim.slope

├

lo.lim.pH(0)

└

up lim.pH(0)

The most important calibration data such as slope and pH(0) must be

within the limits defined here in order to be adopted automatically.

lo.lim.slope

1.0...95.0...999.9 %

up lim.slope

1.0...103.0...999.9 %

lo.lim.pH(0)

0.00...6.40...99.99

up lim.pH(0)

0.00...8.00...99.99

If one of the values of the calibration data lies outside these limits then

at the end of the calibration the following message is displayed:

cal. out of limits

You can accept these calibration data nevertheless with <OK> or you

can reject them with <QUIT>.