3 ph measurement, Ph measurement – Metrohm 827 pH lab User Manual

Page 27

Advertising
background image

4 Short operating tutorial

826/827 pH meter, Instructions for Use

19

6 Possible error messages

•

In case buffer pH 7 has been measured again by mistake, the

following message is displayed:

same buffer

Change the buffer and press <OK>. The calibration is being

continued. The calibration can be canceled with <QUIT>.

•

If the temperature difference is > 2 °C between the buffers,

the following message is displayed:

delta T > 2 °C

Adjust the temperature and continue the calibration with

<OK> or cancel it with <QUIT>

4.3

pH measurement

1 Select print criterion

•

If the obtained measured value is to be printed out directly as

a measuring point report then the required print criterion must

be set (see Section 5.2):

configuration/print crit.: drift

2 Start measurement

•

Immerse the pH electrode in the sample

3 Determine measured value

•

You can read off the current pH value of the sample from the

display. It is stable when the message drifting... is no

longer being displayed (preset drift threshold: 0.028 pH/min)

1

4

.

Print measured value

•

Change to the menu << print >> with the arrow keys <>

or <>.

•

The output of a measuring point report on a printer is started

with <OK>. Due to the print criterion drift the recording of

the measurement and its output takes place only when the

measured value is stable:

1

Further preset drift thresholds:

for potential measurement: U/mV: 1.875 mV/min, for temperature measurement: t/°C: 0.974 °C/min

'mp
827 pH lab 01106 5.827.0010
print date 2005-02-08 15:43:01
pH = 4.612 22.8 °C NTC
==============