3 Installation

IC equipment: Liquid Handling Station

■■■■■■■■

9

Installing the capillary between the trap column and the
peristaltic pump

Required accessories

■

PEEK capillary (6.1831.180)

■

Pressure screw (6.2744.010)

■

Pump tubing with yellow stoppers (6.1823.390)

■

Coupling olive/UNF 10/32 (6.2744.034)

■

Capillary cutter (6.2621.080)

1 Preparing the pump tubing

■

Attach a coupling olive/UNF 10/32 (6.2744.034) to each end of
the pump tubing.
(See the chapter "Installing the peristaltic pump" in the manual
for the Sample Processor.)

2 Cutting the capillary to size

■

Cut off a piece of approx. 70 cm from the long capillary using the
capillary cutter.

3 Mounting the capillary

■

Tighten one end of the capillary to the trap column outlet using a
pressure screw.

■

Tighten the other end of the capillary to the pump tubing inlet
using a pressure screw.

Installing the capillary between the peristaltic pump and
the Liquid Handling Station

Required accessories

■

PEEK capillary (6.1831.180)

■

Pressure screw (6.2744.010)

■

Pump tubing with yellow stoppers (6.1823.390)

■

Liquid Handling Station (6.2841.120)