Metrohm 793 IC Sample Preparation Module User Manual
Page 31

3.3 Operation via 732 IC Detector
793 IC Sample Prep Module/ 8.793.1003 Instructions for Use
27
Rotate SP Module A at an event
In the 732 IC Detector the <EVENT> key can be used to
program a maximum of 4 different events (see section 4.7.3 of
732/733 Manual ). At each event the SP Module A can be brought
into the next position by setting remote output lead 2 to 1 (on,
active, 0 V). This lead must later be returned to the setting 0 (off,
inactive, open) before the SP Module A can be switched to the
next position.
Rotate:
EVENT YY-MM-DD HH:MM:SS
remote
✴
1
✴✴✴✴✴✴
Reset
output lead:
EVENT YY-MM-DD HH:MM:SS
remote
✴
0
✴✴✴✴✴✴
SP Module A position display
STEP
ACTUATOR
The position display 7
777 of the 793 IC Sample Prep Module beside
switch 8
888 shows whether the SP Module A is in the right position:
Lamp on
SP Module A in the right position
Lamp off
SP Module A in the wrong position or blocked
(procedure see section 4.1)