3.15 Mounting the base plate

■■■■■■■■■■■■■■■■■■■■■■

34

■■■■■■■■

855 Robotic Titrosampler

Switch the barcode reader to programming mode and make the following
settings:

1

■

Select the keyboard layout for the desired country (USA, Germany,
France, Spain, Switzerland (German)). This setting must match the
setting in the control software.

■

Make sure that the Ctrl characters (ASCII 00 to 31) are allowed to
be sent.

■

Adjust the settings so that the ASCII character 02 (STX or Ctrl B) is
sent as the first character as "Preamble" or "Prefix Code".

■

Adjust the settings so that the ASCII character 04 (EOT or Ctrl D) is
sent as the last character as "Postamble" or "Record Suffix" or "Post-
fix Code".

■

Exit programming mode.

3.15

Mounting the base plate

If it is needed by an application that a determination is not carried out on
the sample rack but in an external measuring cell, a stand plate can be
mounted. It can be placed on the left or on the right of a tower of the 855
Robotic Titrosampler . Any accessories parts can be placed on the support
rod of the stand plate.

Mounting the base plate

6.2001.070

1

2

3

4

5

6

Figure 28

Mounting the base plate