4.1 Error and special messages
785 DMP Titrino
135
no sequence
No sequence is defined in TIP.
Exit: <CLEAR> and define sequence.
no titration data
No curve can be printed as no data are available.
not possible
Writing functions on the card 6.6036.XXX are not possible.
not valid
A value is not available.
outside
The set fix EP is outside the measuring range.
overrange
The measuring range of
±
2 V has been exceeded. Overrange
replaces the corresponding measured value (pH, U, I or
temperature). If a measured value is in overrange (primary or
secondary measured value), the other (secondary or primary
measured value) can also be unstable.
result out of limits
The result lies outside the limits which were defined in the
method, see page 39.
Exit: Calculate result again or new start.
same buffer
In the calibration sequence the voltage difference between the
first and second buffer is < 6 mV.
Exit: <QUIT> and change buffer or <STOP> (abort
calibration).
sample size out
The sample size is outside the limits which are defined in the
method, see page 18.
Exit: Enter new sample size.
save lines OFF
The function "save lines" is not active although a submethod of
TIP contains assignments to C24 or C25.
Exit: <CLEAR> and switch on "save lines" under <SMPL
DATA> key.
Warning: The results of this line will not be saved.
second TIP call
In TIP no further TIP can be called up as a submethod.
Exit: <CLEAR> and define new sequence.
service is due
The service interval has elapsed. Contact Metrohm service so
that the Titrino can be serviced. This message will appear
each time the Titrino is switched on.
Exit: New start.
silo empty
The silo memory is switched in but empty and a titration has
been started. Corrective action: At least the first 1 silo line
before starting the first titration.
Exit: <CLEAR>.
silo full
The silo memory is full up. Corrective action: If you have filled
less than 255 silo lines, you can create more space by
deleting old methods no longer needed. 1 silo line needs
18...120 bytes.
Exit: <CLEAR>.
stop EP reached
A DET or MET titration was stopped as the stop criterion "stop
EP" was reached.
stop meas.val.reached
A DET or MET titration was stopped as the stop measuring
value pH, U or I was reached.
stop time reached
SET has been stopped as the stop time has been reached.
stop V reached
The determination has been stopped as the stop volume has
been reached.
system error 3
The instrument adjustment data have been overwritten.
Exit: <CLEAR>. Default adjustment data are set. The error