3 Installation

850 Professional IC Anion MSM-HC LP Gradient

■■■■■■■■

23

3.7.2

Positioning and connecting the detector

The instrument is delivered without detector. For information on how to
place and connect the detector, please refer to the manual of the detec-
tor.

3.7.3

Transport locking screws

To avoid damage to the high pressure pump and vacuum pump during
transport, the pumps are secured with transport locking screws .

Remove these transport locking screws before the initial start-up.

Removing transport locking screws

1 Remove all of the transport locking screws with the 6.2621.030 4

mm hexagon key and keep them in a safe place.

Warning

In order to avoid damage to the pump, the transport locking screws
must be remounted each time the instrument undergoes major trans-
port.

3.7.4

Leak sensor

The leak sensor detects escaping liquid which collects in the base tray of
the instrument.

To activate the leak sensor, the leak sensor connector plug (7-2) must be
connected, the instrument switched on and the leak sensor switched to
active in the software.

Connecting the leak sensor

1 Plug the leak sensor connector plug (7-2) into the leak sensor con-

nection socket (7-1) on the rear of the instrument (see Figure 7,
page 24)
.