3.4 Installing the flow-through cell

■■■■■■■■■■■■■■■■■■■■■■

16

■■■■■■■■

944 Professional UV/VIS Detector Vario (2.944.0010)

Installing the flow-through cell

1

2

3

4

1 Loosen and remove the knurled screws (4-3).

2 Remove the cover plate (4-2).

3 Insert the flow-through cell (6.2839.130) such that the opening in

top right corner is aligned with the cylinder screw (4-4) on the cell
block.

4 Screw in the knurled screws again.

NOTE

In order to fix the flow-through cell in the correct position, the
two knurled screws must be tightened symmetrically and with
constant force.

Any tilting, twisting or canting of the flow-through cell influences
the light incidence and thus the measuring results.