3 Installation

850 Professional IC Cation – HP-Gradient

■■■■■■■■

13

3 Connecting the eluent path

Equip an eluent aspiration tubing for each of the two high pres-
sure pumps and connect with the eluent bottles (see Chapter
3.8.1, page 27)
.

Instead of the column, connect the 6.2744.040 coupling with a
6.2744.070 short pressure screw at the end of the pre-installed
column inlet capillary.

Connect the detector inlet capillary with a 6.2744.070 short pres-
sure screw to the .6.2744.040 coupling.i

4 Connecting the sample path

Note

The sample degasser does not have to be connected. We recom-
mend the usage of the sample degasser only if the sample matrix
requires it.

Guide the 6.1803.040 sample aspiration capillary connected to
the sample input of the injection valve out of the instrument
through a capillary feed-through and connect it with the Sample
Processor or

If the sample degasser is used connect the sample aspiration
capillary with a 6.2744.090 long pressure screw to the input of
the sample degasser (see Chapter 3.13, page 39).
Connect a part of the 6.1803.040 sample aspirating capillary with
a 6.2744.090 long pressure screw to the input of the sample
degasser. Guide the other end out of the instrument through a
capillary feed-through and connect it to the Sample Processor.

5 Connecting the instrument

Connect the instrument to the PC with a 6.2151.020 USB cable
(see Chapter 3.16.1, page 46).

Connect instrument to mains supply (see Chapter 3.16, page
46)
.

6 Initial start-up

(see Chapter 4.1, page 50)

Switch on the PC and start MagIC Net.

Switch on the instrument.

Deaerate the high pressure pump (see Chapter 3.10.2, page
35)
.