4 Installation

940 Professional IC Vario ONE (2.940.1100)

■■■■■■■■

23

Rinsing the IC system

CAUTION

The guard column and separation column must not be installed for the
initial start-up.

Make sure that a coupling (6.2744.040) is being used instead of the
columns.

1 Preparing the software

■

Start the MagIC Net computer program.

■

Open the Equilibration tab in MagIC Net: Work-
place

▶ Run ▶ Equilibration.

■

Select (or create) a suitable method.
Also see: MagIC Net tutorial and online help.

2 Preparing the instrument

■

Ensure that the eluent aspiration tubing is immersed in the eluent
and there is enough eluent in the eluent bottle.

■

Switch on the instrument.

MagIC Net detects the instrument and all of its modules.

3 Starting equilibration

■

Start the equilibration in MagIC Net: Workplace

▶ Run ▶ Equi-

libration

▶ Start HW.