76

▪▪▪▪▪▪▪

Noise Summary displayed using the OpQual tab.

The OpQual tab brings up the display shown. This shows results of the Noise Test for each of the four
wavelength regions. These regions are:

400-700nm

700-1100nm

1100-1700nm

1700-2500nm

For each region, results are given for

Peak-to-Peak Noise (P-P)

Root-Mean-Square Noise (RMS)

Bias (A measure of baseline energy changes)

Each of these parameters is described in more detail in the Diagnostics Section of the Vision Manual.
If the test is reported as “Passed” the user may proceed with sample analysis.