2 changing the eluent, 6 servicing the eluent degasser, Changing the eluent – Metrohm 940 Professional IC Vario ONE/Proep 2 User Manual

Page 64: Servicing the eluent degasser

Advertising
background image

5.6 Servicing the eluent degasser

■■■■■■■■■■■■■■■■■■■■■■

56

■■■■■■■■

940 Professional IC Vario ONE/Prep 2 (2.940.1120)

The composition of the eluent plays a critical role in chromatographic ana-
lysis:

Concentration

An increase in the concentration generally leads
to shorter retention times and faster separation,
but also to a higher background conductivity sig-
nal.

pH

pH changes lead to shifts in dissociation equili-
bria and thus to changes in retention times.

Organic solvents

Adding organic solvents (such as methanol, ace-
tone or acetonitrile) to a watery eluent generally
speeds up lipophilic ions.

5.5.2

Changing the eluent

Ensure that no precipitates can form when changing the eluent. Immedi-
ately successive solutions must be miscible. If the system has to be rinsed
with organic solvents, several solvents with rising or falling lipophilicity
must be used.

NOTE

To change the eluent, remove the separation column and connect the
capillaries using a coupling (6.2744.040) and two pressure screws
(6.2744.070).

5.6

Servicing the eluent degasser

The pieces of connection tubing are connected to the eluent degasser
with clamping screws.

Proceed as follows if you have to loosen and reconnect the connection
tubing at the eluent degasser:

Accessories

For this step you need the following accessories:

You can find these parts in the accessory kit: Vario/Flex Basic (6.5000.000)

Wrench (6.2621.050)

Loosening the connecting tubing

1

Loosen the clamping screws with the wrench.

Unscrew the clamping screws by hand and pull them out of the
connector.

Advertising