7 sensor, Sensor – Metrohm 840 PC Control 6.0 / Touch Control User Manual

Page 22

Advertising
background image

2.1 Coulometric Karl Fischer titrations (KFC)

■■■■■■■■■■■■■■■■■■■■■■

16

■■■■■■■■

Control device

Selection of the control device from the list of devices. Only those devices
are displayed which are able to carry out the command. The control devi-
ces do not have to be connected. This allows to transfer methods more
easily from one system to another.

Selection

Selection of configured control devices

2.1.7

Sensor

Under [Sensor], the parameters for the sensor are edited.

Sensor

Selection of the sensor from the sensor list. The selection depends on the
measuring mode. Sensors are defined under System

▶ Sensors. You can

also enter a sensor name which is not contained in the sensor list. When a
determination is started there is a check whether the sensor is contained
in the sensor list.

Selection

Selection of configured sensors

I(pol)

The polarization current is the current that is applied to a polarizable elec-
trode during the voltametric measurement.

Selection

5 µA | 10 µA | 20 µA | 30 µA

Default value

10 µA

Electrode test

on | off (Default value: off)

For polarizable electrodes and the generator electrode, an electrode test
can be carried out. The following will be tested:

■

Polarizable electrodes

– Is the electrode connected?
– Is there a short circuit?

The electrode test is carried out when this command is started.

■

Generator electrode:

– Is the electrode connected?
– Is the generator system defective?
– Is the resistance of the sample solution very high?

The electrode test is carried out continuously during the execution of
this command.

Temp. measurement

Type of temperature measurement.