3 software settings, Software settings – Metrohm 797 VA Computrace User Manual
Page 59

2.9
Connection of 863 Compact Autosampler
797 VA Computrace / Hardware-Manual 8.797.8001EN
51
2.9.3 Software
settings
Before putting the 797 VA Computrace Stand with the 863 Compact Autosampler in-
to operation, the following settings have to be made in the "797 VA Computrace
Software" software program:
1 Set dosing parameters
• Click on
797 VA COMPUTRACE
/ Settings / General settings
and select the
Dosi-
nos
tab (if the Dosino is connected to the 846 Dosing Interface: the
Dosing
Interface
tab) of the window
General settings
and enter the following con-
figuration:
• Click on
to display the currently connected Dosing devices.
• Set parameters of used Dosing devices.
• Define the addition or predose solutions for the desired method in the
DOSINOS
window (procedure see Software Manual, section 5.2).
• Fill the Dosing- or Exchange units of the Dosinos/Dosimats with the desired
solutions and make sure that there are no gas bubbles in the glass cylinders
using the
button in the
DOSINO CONTROL
window (see section
7.2, Software Manual).
• Note: If you choose Port 3 for menu item
Prep / Empty via port
, you must
install an FEP Tubing Connection 6.1805.XXX from Port 3 to a waste con-
tainer.