4 appendix, 1 technical data – Metrohm 797 VA Computrace User Manual
Page 90

4 Appendix
797 VA Computrace / Hardware-Manual 8.797.8001EN
82
4 Appendix
4.1 Technical
data
Subject to changes!
The listed technical data apply to an ambient tempera-
ture of 25°C.
Brief characterization
PC-controlled system for voltammetry, set chemical workplace with potentiostat
and measuring amplifier.
With Multi-Mode Electrode or rotating disk electrode (RDE).
Tilt-back measuring arm, integrated drip pan.
Current measurement techniques
DC
Direct Current
NP
Normal Pulse
DP
Differential Pulse
SqW
Square Wave (10 ... 2000 Hz)
AC1
Phase-sensitive Alternating Current 1
st
harmonic
(1 … 250 Hz)
AC2
Phase-sensitive Alternating Current 2
nd
harmonic
(1 … 250 Hz)
PSA
Potentiometric Stripping Analysis (chronopotentiometry)
with chemical oxidation/reduction
CCPSA
PSA with Constant Current oxidation/reduction
CV
Cyclic Voltammetry (digital ramp)
CVS
Cyclic Voltammetric Stripping
CPVS
Cyclic Pulse Voltammetric Stripping (Chronoamperometry)
Potentiostat
Output voltage (AE)
± 12 V
Output current (AE)
± 80 mA
Sweep voltage range
± 5 V
Voltage resolution
150 μV
Input impedance(RE)
R ≥ 1·10
10
Ω
Input Bias Current (RE)
± 5 pA
Noise
typ. 200 pA
Sweep rate
with voltage step 10 mV
CV,CVS: 0 … 36.7 V/s
SQW, DC: 0 … 20 V/s
DP, NP: 0 … 0.5 V/s
AC1, AC2: 0 … 22 mV/s