3 Safety
797 VA Computrace / Hardware-Manual 8.797.8001EN
76
3.2 Change
fuses
The fuses of the 797 VA Computrace are placed between the mains switch 14 and
the mains connection plug 16 under the Fuse cover. To change them, proceed as fol-
lows:
1 Disconnect mains cable
Disconnect mains cable from mains connection plug 16 of the
797 VA Computrace.
2 Remove fuse cover
Using a screwdriver, lever out fuse cover 15 forwards until it
opens.
3 Check and change fuses if necessary
Carefully take the fuses installed for the desired mains voltage out
of the fuse holder and check its specifications:
100…240 V 1.6 AHT Metrohm-Nr.
U.600.0018
Change fuses if necessary and reinsert in fuse holder.
4 Install fuse holder
Reinsert fuse holders in the instrument (the arrows printed on the
holders must point in the same direction as the arrows on the
inside of fuse cover 15).
5 Install fuse cover
Push in fuse cover 15 firmly until it clicks into place.
6 Connect mains cable
Plug mains cable into mains connection plug 16.
3.3 Cabinet
temperature
The cabinet of the 797 VA Computrace stand locally heats up to 45°C or more.