Metrohm 872 Extension Module Suppression User Manual

Page 42

Advertising
background image

5.4 Suppressor/SPM

■■■■■■■■■■■■■■■■■■■■■■

34

■■■■■■■■

872 Extension Module Suppression

Regenerating the suppressor

1 Disconnecting the suppressor/SPM from the IC system

■

Disconnect the capillaries of the suppressor/SPM labeled with
regenerant and rinsing solution from the IC system.

2 Connecting the suppressor/SPM to the high pressure pump

■

Connect the inlet capillary for the regeneration solution (labeled
with regenerant) to the outlet of the high pressure pump with the
aid of a coupling (6.2744.040)

3 Regenerating the suppressor/SPM

■

Regenerate the first suppressor unit/neutralization unit for about
15 minutes.

■

In the software, use the Step command to switch to the second
suppressor unit/neutralization unit and regenerate this for about
15 minutes.

■

In the software, use the Step command to switch to the third
suppressor unit/neutralization unit and regenerate this for about
15 minutes.

4 Rinsing the suppressor

After regeneration, the three suppressor units/neutralization units
must be rinsed with degassed ultra pure water for about 15 minutes.

■

Remove the inlet capillary for the regeneration solution (labeled
with regenerant) from the outlet of the high pressure pump.

■

Connect the inlet capillary for the rinsing solution (labeled with
rinsing solution) to the outlet of the high pressure pump with the
aid of a coupling (6.2744.040).

■

Rinse the first suppressor unit/neutralization unit with degassed
ultra pure water for about 15 minutes.

■

In the software, use the Step command to switch to the second
suppressor unit/neutralization unit and rinse this for about 15
minutes.

■

In the software, use the Step command to switch to the third
suppressor unit/neutralization unit and rinse this for about 15
minutes.

5 Connect the suppressor to the IC system

■

Reconnect the capillaries labeled regenerant and rinsing solution
to the IC system.