4.1 Technical data
771 IC Compact Interface
17
4 Appendix
4.1 Technical
data
Mains connection
Voltage
100...240 V
Frequency
50...60 Hz
Power consumption
6 W
Fuse
2
×
1 ATH (to be replaced by Metrohm Service
only using the same type)
Additional electronic overload protection
PC interface (RS232)
Connector
Dsub connector 9 pin (male)
1
5
9
6
Baud rate
1'200
…
115'200
Data bits
8
Stop bits
1
Handshake
SWChar, none
Parity
none, even, odd
Pin assignment
Pin 1,4,6: internally connected
Pin 2:
RxD (Received Data)
Pin 3:
TxD (Transmitted Data)
Pin 5:
GND (Signal Ground)
Pin 7,8:
internally connected
Analog signal interface
Function
2 potential-free analog signal inputs per system.
Connector
Dsub connector 9 pin (female)
1
5
6
9
Input voltage range
±
2.5 V
Input amplification
1, 2, 4, 8, 16, 32, 64 (selectable)
Sampling rate
10, 20, 30, 50, 60 measuring points/s
Resolution
24 Bit (1 LSP = 0.298
µ