3.1 Equipping the eluent bottle

■■■■■■■■■■■■■■■■■■■■■■

4

■■■■■■■■

IC equipment: Additional eluent for 941

3 Installation

Proceed as follows to install the IC equipment: Additional eluent for 941:

■

Equip the eluent bottle.

■

Equip the concentrate bottle.

■

Equip the ultrapure water container.

■

Assemble the 800 Dosino and connect all tubings.

■

Plug level sensor cable into the 941 Eluent Production Module.

3.1

Equipping the eluent bottle

For the automated eluent production, the eluent bottle has to be
equipped with a bottle cap for level sensor and a sensor. The sensor
choice depends on which fill level is to be monitored: bottle full (FULL) or
bottle empty (EMPTY).

Inserting the measuring sensor

Required accessories

■

A bottle cap for level sensor (6.1626.000)

■

FULL sensor (6.2769.000) for anion eluents or EMPTY sensor
(6.2769.1X0) for cation eluents

Figure 2

Bottle cap for level sensor – Openings

1

M6 opening

2

M6 opening

3

M8 opening

4

Opening for level sensor cable

5

Opening for level sensor cable

1 Remove the stoppers from the bottle cap.

2 Place the bottle cap on a surface with the opening facing up.