3.1 Equipping the eluent bottle

■■■■■■■■■■■■■■■■■■■■■■

6

■■■■■■■■

IC equipment: Additional eluent for 941

1 Mounting the adsorber tube

Fill the adsorber tube with adsorber material and cotton and insert it
into the M6 opening (2-1).

2 Connecting the FEP tubing

Fasten the tubing connection to the Dosino to the M6 opening (2-2).

■

Insert the FEP aspiration tubing into the M6 opening.

■

Screw the FEP tubing into the M6 opening.

3 Connecting the eluent aspiration tubing

Fasten the capillary connection to the ion chromatograph to the M8
opening (2-3).

■

Start by pushing the M8 tubing nipple over the end of the aspira-
tion tubing, followed by the O-ring.

■

Guide the eluent aspiration tubing through the M8 opening of
the bottle cap and temporarily secure it in place with the tubing
nipple.

■

Attach the tubing weighting and the aspiration filter to the end of
the eluent aspiration tubing, see the manual of the ion chro-
matograph
.

4 Attaching the bottle cap

■

Screw the bottle cap onto the eluent bottle.

■

Loosen the tubing nipple a little.

■

Push the eluent aspiration tubing so far into the eluent bottle that
the aspiration filter is at the bottom of the bottle.

■

Tighten the tubing nipple again.

5 Connecting the level sensor cable

Connect the two plugs of the level sensor cable to the two sockets
(2-4) and (2-5).