Metrohm Additional eluent for 941 User Manual

Page 17

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

IC equipment: Additional eluent for 941

■■■■■■■■

11

3 Connecting the Dosino to the ion chromatograph

NOTE

The 941 Eluent Production Module must be switched off when
the Dosino is being plugged to the MSB connector.

■

Check whether the 941 Eluent Production Module is switched on.
If so, switch off the instrument.

■

Plug the Dosino cable into one of the 941 Eluent Production
Module's MSB sockets.

Installing the FEP tubings on the Dosino

Accessories

For this step you need the following accessories:

■

800 Dosino (2.800.0010) with 807 Dosing Unit 50 mL (6.3032.250)
attached

■

The three FEP tubings connected to the bottles (6.1805.530),
one additional FEP tubing (6.1805.530)

■

Adapter Dosino Port 4 (6.1808.280)

■

Adapter UNF / M6 (6.2744.080)

1 Connecting the lateral ports

1

Concentrate

2

Port 1

3

Port 3

4

Eluent

■

Screw the concentration bottle's FEP tubing onto Port 1 of the
Dosino.

■

Screw the eluent bottle's FEP tubing onto Port 3 of the Dosino.