3 Installation

IC equipment: Additional eluent for 941

■■■■■■■■

9

■

Adapter for adsorber tube (6.1624.000)

■

M6 threaded stopper (6.1446.040), included in the accessories for the
eluent bottle cap (6.1602.160)

Figure 6

Installing the ultrapure water canister cap without level sen-
sor

1 Connecting the FEP tubing

Fasten the tubing connection to the Dosino to the M6 opening.

■

Insert the FEP aspiration tubing into the M6 opening.

■

Screw the FEP tubing into the M6 opening.

2 Sealing the M8 opening

■

Screw the M8 threaded stopper into the M8 opening.

3 Mounting the adsorber tube

■

Fill the adsorber tube with soda lime and cotton.

■

Fasten the adsorber tube in the large opening with the adapter.