4 eluent, 1 connecting eluent bottle, Eluent – Metrohm 872 Extension Module User Manual

Page 31: Connecting eluent bottle, Figure 11, Installation diagram, Figure 11 installation diagram

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Installation

872 Extension Module IC Module

■■■■■■■■

23

Eluent

Degasser

In

Out

Sample

Eluent

Standard

1

2

3

4

5

6

1

2

3

4

Figure 11

Installation diagram

1

Eluent aspiration tubing 6.1834.080
Connected to the eluent degasser. Connect
the other end to the eluent bottle.

2

Eluent connection capillary 6.1831.010

3

Sample aspirating capillary 6.1803.040

4

Sample outlet capillary 6.1803.040

The following chapters describe the individual installation steps in detail.

4.4

Eluent

4.4.1

Connecting eluent bottle

The eluent is aspirated out of the eluent bottle via the eluent aspiration
tubing (12-1).

The eluent aspiration tubing is connected to the eluent degasser (see
Chapter 4.5, page 27)
. The tubing must be threaded through a suitable
capillary feed-through of the instrument before the other end can be
equipped.

You will require the parts from the following accessories for equipping the
eluent aspiration tubing:

6.1602.160 Eluent bottle attachment GL 45

6.2744.210 tubing adapter for aspiration filter

6.2821.090 aspiration filter

To equip the eluent aspiration tubing proceed as follows:

Assembling eluent aspiration tubing

1 Guide the free end of the eluent aspiration tubing (12-1) out of the

instrument through a suitable capillary feed-through.