Metrohm 940 Professional IC Vario ONE/SeS/PP/Prep 3 User Manual
Page 66

3.21 Initial start-up
■■■■■■■■■■■■■■■■■■■■■■
56
■■■■■■■■
940 Professional IC Vario ONE/SeS/PP/Prep 3 (2.940.1530)
4 Deaerating the high-pressure pump
■
Push the end of the purging needle (6.2816.040) over the end of
the deaerating capillary on the purge valve.
■
Insert the syringe (6.2816.020) in the Luer connector of the purg-
ing needle.
■
Open the purge valve using the rotary knob (approx. ½ turn).
■
Switch on the high-pressure pump in MagIC Net.
■
Use the syringe to aspirate eluent until there are no more air bub-
bles in the aspirated eluent.
■
Switch off the high-pressure pump in MagIC Net.
■
Seal the purge valve using the rotary knob.
■
Remove the syringe from the purging needle.
■
Pull the purging needle out of the deaerating capillary.
5 Setting the contact pressure of the peristaltic pump
■
6 Rinsing the instrument without columns
■
Rinse the instrument (without columns) with eluent for 5 minutes.