3.4 Removing transport locking screws
■■■■■■■■■■■■■■■■■■■■■■
18
■■■■■■■■
940 Professional IC Vario ONE/ChS/PP (2.940.1300)
3.4
Removing transport locking screws
To avoid damage to the drives for the high-pressure pump and the vac-
uum pump during transport, the pumps are secured with transport lock-
ing screws. These are located at the rear of the instrument and labeled
with Transport security screws.
Remove these transport locking screws before the initial start-up.
Accessories
For this step you need:
■
Hex key 4 mm (6.2621.030)
Remove the transport locking screws
Figure 6
Removing the transport locking screws
1
Transport locking screws
For the vacuum pump.
2
Transport locking screws
For the high-pressure pump.
3
Transport locking screws
For an additional high-pressure pump in the
bottom drawer.
1 Remove all of the transport locking screws with the hex key.