2.2 Mains connection

782 IC Column Thermostat

9

88

99

12

12

230

77

115

Fig. 3: Setting the mains voltage

2.2.2

Fuses

Two fuses 0.5 A/slow-blow for 100

120 V or 0.25 A/slow-blow for

220

240 V are installed in the fuse holder 77 of the 782 IC Column

Thermostat as standard.

Ensure that the instrument is never put into operation with fuses of

another type, otherwise there is danger of fire!

For checking or changing fuses, proceed as described in section 2.2.1.

2.2.3

Mains cable and mains connection

Mains cable

The instrument is supplied with one of three mains cables

6.2122.020 with plug SEV 12 (Switzerland,

)

6.2122.040 with plug CEE(7), VII (Germany,

)

6.2133.070 with plug NEMA 5-15 (USA,

)

which are three-cored and fitted with a plug with an earthing pin. If a

different plug has to be fitted, the yellow/green lead (IEC standard)

must be connected to protective earth (protection class 1).

Any break in the earthing inside or outside the instrument can make it

a hazard!

Mains connection

Plug the mains cable into mains connection plug 12

12 of the 782 IC Col-

umn Thermostat (see Fig. 3).

2.2.4

Switching the instrument on/off

The 782 IC Column Thermostat is switched on and off using mains

switch 99. When the instrument is switched on the two temperature dis-

plays 55 and 66 light up.

100 – 120 V

220 – 240 V

77 Fuse holder

88 Voltage selection

insert with display

of voltage

99 Mains switch

12

12 Mains connection

plug