Metrohm NIRS XDS Process Analyzer – DirectLight/NonContact User Manual

Page 125

Advertising
background image

▪▪▪▪▪▪▪

123

11.

Install the lamp cover back onto the lamp

housing. The widest part faces the user. Do
not try to install in the wrong orientation.

12.

Tighten the four (4) captive thumbscrews

on the lamp cover.

Do not use tools to tighten these screws, as
the heat cycling of the instrument may
make them hard to remove. Hand-tight is
sufficient.

13.

Close the instrument. Apply power to

warm up the instrument.

14.

Enter Vision Software and connect to the instrument.

15.

Run Wavelength Linearization.

16.

Run a new Reference Standardization.

17.

Run Instrument Calibration.

18.

Run Performance Test.