Application Bulletin 353

Installation Instructions of TitrIC pro I

Page 9 of 14

HCl

Port Length

Diameter

Dosing Port
Prep/Empty

Dosing Port 1

-

-

Dosing Port 1

Port 1

40 cm

2.0 mm

Dosing Port 2

Port 3

0 cm

0.0 mm

Fill Port

Port 2

25 cm

2.0 mm

Special Port

Port 4

0 cm

0.0 mm

EDTA

Port Length

Diameter

Dosing Port
Prep/Empty

Dosing Port 1

-

-

Dosing Port 1

Port 1

60 cm

2.0 mm

Dosing Port 2

Port 3

0 cm

0.0 mm

Fill Port

Port 2

25 cm

2.0 mm

Special Port

Port 4

0 cm

0.0 mm

TRIS/acetyacetone

Port Length

Diameter

Dosing Port
Prep/Empty

Dosing Port 1

-

-

Dosing Port 1

Port 1

60 cm

2.0 mm

Dosing Port 2

Port 3

0 cm

0.0 mm

Fill Port

Port 2

25 cm

2.0 mm

Special Port

Port 4

0 cm

0.0 mm

Fig.: Configuration table of the Dosing Units

6.6 Electrodes

pH electrode: Under ‘

Configuration/Sensors’

click on

‘iAquatrode Plus’,

check the information and click on

<Ok>

.

Ca ISE electrode: Under ‘

Configuration/Sensors’

click

on ‘

Edit/New/ISE electrode

and enter under ‘

Sensor

name’ Ca-ISE

, type in the desired sensor information

and click on

<Ok>

.

Conductivity sensor: Under ‘

Configuration/Sensors’

click

on ‘

Edit/New/other sensor’

and enter under ‘

Sensor

name’ Conductivity sensor

type in the desired sensor

information and click on

<Ok>

.

6.7 Methods

The following methods (.mmet files) are provided for
tiamo

TM

:

TitrIC pro I:

This method is the standard method used. It

does the complete titration analysis and sends the

results to MagIC Net. It is used in combination with

TitrIC pro I

method in MagIC Net.

In tiamo it is typically used as a single determination in
the autorepeat mode. In other words, it is always ready
to do a titration as soon MagIC requires it. To set it into
autorepeat go to the Run window, right-mouse-click and
select ‘

Properties Run window’

. Under Process select

Autorepeat determination’

. Close the window with

<Ok>

.

TitrIC pro I_pH calibration manual:

Here the iAquatrode

is taken out of the titration vessel and placed into the
buffer solutions. Start this method only from tiamo

TM

!

TitrIC pro I_titer determination of HCl:

This method is

used for the titer determination. It has to be used in
combination with the MagIC Net method

TitrIC pro I_titer

determination of HCl

.

TitrIC pro I_titer determination of EDTA:

This method is

used for the titer determination. It has to be used in
combination with the MagIC Net method

TitrIC pro I_titer

determination of EDTA

.

TitrIC pro I_conductivity calibration:

This method is used

for calibration of the conductivity sensor. It has to be
used in combination with the MagIC Net method

TitrIC

pro I_calibration conductivity

.

Import: Click on

Method

at the left, then select under

File /

Method manager… / Edit / Import…

Import the methods

related to TitrIC pro I also from the MagIC Net CD. You will
find the methods in the folder example/TitrIC of the CD.
Note!
Alterations of these methods should only be carried out by a
person at the administrator level and who is thoroughly
familiar with tiamo

TM

.

Shut down tiamo

TM

6.8 Setup of RS232 communication between MagIC Net

and tiamo

TM

The communication, e.g. result transfer between MagIC Net
and tiamo

TM

is done via RS232. MagIC Net is the master in

this setup. To establish the communication follow the steps
below:

Make sure MagIC Net and tiamo

TM

software are closed. If

the PC has two built-in RS232 ports you can connect the
RS232 cable (6.2134.040) directly between the two ports. If
the PC does not have the built-in RS232 ports please
connect two USB/RS232 Converters (6.2148.050) to two
USB ports. You can e.g. use the USB hubs of the IC

Advertising