3 Installation

930 Compact IC Flex ChS/Deg (2.930.1260)

■■■■■■■■

43

MagIC Net detects the instrument and all of its modules.

3 Starting equilibration

Start the equilibration in MagIC Net: Workplace

Run Equi-

libration

Start HW.

4 Deaerating the high-pressure pump

Push the end of the purging needle (6.2816.040) over the end of
the deaerating capillary on the purge valve.

Insert the syringe (6.2816.020) in the Luer connector of the purg-
ing needle.

Open the purge valve using the rotary knob (approx. ½ turn).

Switch on the high-pressure pump in MagIC Net.

Use the syringe to aspirate eluent until there are no more air bub-
bles in the aspirated eluent.

Switch off the high-pressure pump in MagIC Net.

Seal the purge valve using the rotary knob.

Remove the syringe from the purging needle.

Pull the purging needle out of the deaerating capillary.

5 Rinsing the instrument without columns

Rinse the instrument (without columns) with eluent for 5 minutes.