6 Devices
IC Net 2.1
64
Remote lines after power on
The remote output lines 1}8 are set to this
values after power on or a manual stop with
Shutdown hardware
.
Selection:
0
,
1
Thermostat
Operating temperature of the conductivity
cell.
Selection:
25
,
30
,
35
,
40
,
45 °C
,
off
Thermostatting functions only if the ambient tempera-
ture is at least 5 °C lower than the operating tempera-
ture. It normally takes 30}60 min after power on until
a temperature stability of r 0.01 °C is attained.
Flow correction
Factor for correction of the difference be-
tween displayed and actual flow rate of the
high-pressure pump.
Range:
0.9
...
1.09
The flow correction is determined by meas-
urement of the actual flow rate with the aid
of a measuring cylinder as follows:
Flow correction
=
Displayed flow rate
Measured flow rate
Cell constant
Cell constant of the conductivity cell for cor-
rect display of the absolute conductivity.
Enter the cell constant printed on the detec-
tor block into this field.
Range:
0.1
...
99.9 /cm
For a precise determination of the cell con-
stant, pump a calibration solution of known
conductivity through the IC system, observe
the displayed conductivity and change the
cell constant until the correct conductivity
value is displayed.
<Break times>
Possibility for changing the break times for
injection valve
Valve
and suppressor mod-
ule
Suppressor
.
These values should only be changed af-
ter consulting the Metrohm Service.