6.5 817 Bioscan

IC Net 2.1

73

Method data

Cell potential

Definition of start and end potential for a
user-defined potential sweep, which is ap-
plied to the amperometric detector cell.

E1

Start potential for sweep.
Range:

-2 ... +2 V

.

E2

End potential for sweep.
Range:

-2 ... +2 V

.

Scan cycle

Definition of scan cycle:

full

double

scan

E1 ... E2 ... E1

half

single

scan

E1 ... E2

Selection:

full, half

Scan rate

Definition of scan rate in mV/s.
Selection:

1, 2, 5, 10, 20, 50 mV/s

Oven

Switching on/off column oven and setting
operating temperature.
Range:

15 ... 60 °C

Range

Selection of measuring range:

pA

10 ... 5000 pA

.

nA

0.1 ... 50 nA

.

µA

0.01 ... 5 µA

.

Full

scale

The full scale range (operating range) sets
the desired sensitivity for the analog output.
The possible values of the full scale range
depend on the preset measuring

Range

.

Offset

Offset of the zero point for the analog out-
put signal.
Selection:

-50 ... +50 %

(in steps of 10 %)

Actual

E1, E2

Display of current start and end potentials
in V.

Scan cycle

Display of current scan cycle mode.

Scan rate

Display of current scan rate in mV/s.

Oven

Display of current oven state.

Range

Display of current measuring range.

Full

scale

Display of current full scale range.

Offset

Display of current offset.

<Set>

Send current parameters immediately to the
817 Bioscan. Parameters are not stored in
the system file (

*.smt

) as long as the file is

not saved.