Preparing the determination of the temperature correction

The figure shows in detail how the accessory parts are to be assembled
for the determination of the temperature correction. Proceed as follows:

1 Preparing the reaction vessel cover

■

Mount the air tube on the reaction vessel cover.

■

Clamp the first spacer at a distance of approx. 12 cm from the
lower end onto the air tube.

■

Clamp the second spacer at a distance of approx. 5 cm from the
lower end onto the air tube.

■

Insert the temperature sensor from above into the temperature
sensor opening and fasten it in the corresponding openings of the
spacers.

2 Preparing the reaction vessel

■

Fill the reaction vessel with 6 g silicone oil (6.2326.000).

■

Place the reaction vessel cover with the temperature sensor on
the reaction vessel.

■

Push the temperature sensor all the way down (the sensor must
touch the vessel base).

3 Inserting and connecting the reaction vessel

■

Insert the reaction vessel with the mounted reaction vessel cover
in the measuring position 2 or 3 of the required heating block.

■

Screw one end of the FEP tubing 250 mm to the M8/M6 thread
adapter of the reaction vessel cover.

■

Screw the other end of the FEP tubing to the corresponding con-
nector of the 892 Professional Rancimat.

■

Connect the temperature sensor to the Pt100 connector (2-14)
on the rear of the 892 Professional Rancimat.

NOTE

You will find detailed information regarding the StabNet computer
program in the Tutorial.